Evaluation Of Doses Delivered During Ct Examinations By Different Scanners For Purposes Of Intercomparison And Dose Optimization.
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
University of Ghana
Abstract
This research study was aimed at performing dosimetry intercomparison on different CT
scanners in the diagnostic radiology departments of Korle-Bu Teaching Hospital (KBTH),
Sweden Ghana Medical Center (SGMC) and Global Medical and Imaging Center (GMIC).Using
the standard body phantom and integrated ion chamber technique volume computed tomography
dose index (CTDIvol) and Dose-Length Product (DLPs) within the phantom were evaluated. The
ion chamber technique was applied to two 16 slice Siemens and one Toshiba Aquilion one CT
scanners. CTDIvol and DLP values for the standard body polymethyl methacrylate ( PMMA)
phantom were estimated and comparison made with corresponding console displayed values for
accuracy and also to deduce a suitable method for optimization of patients and occupationally
exposed worker doses. Effective doses were also calculated. An intra and inter institutional
comparison of measured doses and console displayed doses were performed. Chest protocol at
Automatic Exposure Control ( AEC) was applied during the scanning of the phantom. Estimated
CTDIvol values (mGy) were 17mGy, 24mGy and 13.1mGy for SGMC, GMIC and KBTH
respectively. These values deviated from the console displayed values by 24.1%, 22.9%
and31.3% respectively. Similarly, estimated DLP values (mGy.cm) were
675mGy.cm,944mGy.cm and 419mGy.cm for SGMC, GMIC and KBTH respectively deviating
from the console displayed values by 24.1%, 24.2% and 29% respectively. In terms of effective
doses (E), the calculated E (mSv) values were 9.45mSv, 13.2mSv and 5.87mSv estimated from
the DLPs from SGMC, GMIC and KBTH respectively using K ,the anatomy-specific dose
coefficient expressing effective dose normalized to DLP in a standard CT dosimetry phantom of
0.014 mSv mGy-1 cm-1.The estimated doses were compared to other selected international
Dose Reference Levels ( DRLs) and were within range.
Description
Thesis (MPhil)