Development Of Optimised Protocols For Paediatric Whole-Body Computed Tomography Examinations: A Figure-Of-Merit Approach.
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
BMC Medical Imaging
Abstract
Background Whole-body computed tomography (WBCT) technique involves higher radiation doses, resulting
in increased stochastic health risks, especially when used for paediatric patients. Hence, protocol optimisation is
recommended to maximise its benefit-risk ratio, with several available strategies to achieve it. This study used the
figure-of-merit (FOM) approach to develop optimised protocols for paediatric WBCT examinations. The rationale is to
enhance diagnostic accuracy while minimising radiation exposure, ultimately improving patient safety and outcomes
for paediatric patients undergoing WBCT.
Methods Newborn and child (5-year-old) anthropomorphic phantoms were scanned with different scan protocols
and evaluated for dose and image quality using the CT-Expo and ImageJ programs, respectively. The protocols
for trunk examinations were divided into arterialphase-only and combined arterial and venous phase to develop
appropriate protocols based on patients’ initial focused assessment with sonography for trauma (FAST) results. The
protocols with the highest FOMs were selected as the best optimised protocols.
Results The estimated WBCT ED (WBED) for the optimised protocols ranged from 2.6 mSv to 5.8 mSv with estimated
FOM of 3.83 to 59.35. The mean effective doses (EDs) for newborn and child arterial phase-only protocols were not
significantly lower than the combined arterial and venous phase protocols (p=0.069, 0.082 respectively), while the
mean signal-to-noise ratio of the combined phase protocols for newborn and child phantoms were insignificantly
higher than the single-phase protocol (p=0.089, 0.662 respectively).
Conclusion The estimated EDWB from this study were lower than established values. The selected protocols are
recommended for newborns and children (1–7 years) using the GE Revolution CT and Siemens SOMATOM Definition
Edge CT scanners.
Description
Research Article
Citation
Ago, J. L., Inkoom, S., Ohene-Botwe, B., Larsen, A., & Berg, I. S. (2025). Development of optimised protocols for paediatric whole-body computed tomography examinations: a figure-of-merit approach. BMC Medical Imaging, 25(1), 125.