Errors In Manual Radiotherapy Treatment Procedures And Their Evolution In A Low Resource Setting: Uganda’s Experience

dc.contributor.authorKomakech, I.
dc.contributor.authorTagoe, S.N.A.
dc.contributor.authorOrem, J.
dc.contributor.authoret al.
dc.date.accessioned2024-02-12T15:13:05Z
dc.date.available2024-02-12T15:13:05Z
dc.date.issued2023
dc.descriptionResearch Articleen_US
dc.description.abstractPurpose: In Uganda, two-dimensional (2D) radiotherapy treatments have been in use since the establishment of radiotherapy in 1995. Preliminary investigations of treatment records in November 2019 showed evidence of gaps requiring urgent attention. The purpose of this study was to improve the safety of the treatments. Methods: Records of 1164 patients treated in 1387 courses (1412 sites) on Cobalt-60 units were reviewed to determine the frequency and dosimetric implications of events that occurred at different steps of the radiotherapy process. The results were presented and discussed with the different professionals for learning purposes. Results: Most common dosimetric events were omission of block tray, bolus and couch transmission factors in time calculations, incorrect field sizes and depths, wrong beam weighting, independent calculations and prescription doses contributing 28.6%, 10.1%, 6.0%,11.9 %, 10.1%, 5.4%, 4.8%, and 8.9% to the 168 observed errors. Comparison of the calculated treatment doses with the prescribed doses showed that 88 % of the 1412 sites were treated with radiation doses with an accuracy of ± 5%. However, an analysis of the evolution along the years demonstrated an improvement from 82.8% in 2018 to 86.1% in 2019 and 93.2% in 2020. Most common procedural events were incomplete setup instructions and missing patient data in the record and verify system of the Co-60 units for 57% and 61.1% of the 1164 patients, respectively. Conclusions: Opportunities for improving safety in the delivery of radiotherapy treatments were identified. Learning from these past errors should raise awareness within the team, leading to safer treatments.en_US
dc.identifier.otherhttps://doi.org/10.1016/j.ejmp.2024.103212
dc.identifier.urihttp://ugspace.ug.edu.gh:8080/handle/123456789/41154
dc.language.isoenen_US
dc.publisherElsevier Ltd.en_US
dc.subjectRadiotherapy quality and safetyen_US
dc.subjectQuality managementen_US
dc.subjectLearning from eventsen_US
dc.titleErrors In Manual Radiotherapy Treatment Procedures And Their Evolution In A Low Resource Setting: Uganda’s Experienceen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Errors In Manual Radiotherapy Treatment Procedures And Their Evolution In A Low Resource Setting Uganda’s Experience.pdf
Size:
1.24 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: