Construction and pre-evaluation of an in-house cylindrical ionization chamber fabricated from locally available material

dc.contributor.authorTagoe, S.N.A.
dc.contributor.authorChaphuka, C.D.
dc.contributor.authorHasford, F.
dc.date.accessioned2023-01-25T10:03:50Z
dc.date.available2023-01-25T10:03:50Z
dc.date.issued2022
dc.descriptionResearch Articleen_US
dc.description.abstractIntroduction: The objectives of this study were to construct a very robust in-house cylindrical ionization chamber from locally available materials to minimize cost, and to assess its suitability for use in a clinical setting. Materials and Methods: The entire body of the constructed IC was composed of Perspex (PMMA). Other components of the IC were made from locally available materials, such as paper and discarded items. The in-house IC was made waterproof by passing the triaxial cable connecting its various electrodes through a plastic tube which once served as a drainage tube of a urine bag. This connection was made such that the chamber was vented to the environment. The completed in-house IC was evaluated for: polarity effect, ion recombination, ion collection efficiency, stability, dose linearity, stem effect, leakage current, angular, dose rate and energy dependences. Results: Although the pre-evaluation results confirmed that the in-house IC satisfied the stipulated international standards for ICs, there was a need to enhance the stem effect and leakage current characteristics of the IC. The in-house IC was found to have an absorbed dose to water calibration coefficient of 4.475 x 107 Gy/C (uncertainty of 1.6%) for cobalt 60 through a cross-calibration with a commercial 0.6 cc cylindrical IC with traceability to the Germany National Dosimetry Laboratory. Using a Jaffé diagram, the in-house IC was also found to have a recombination correction factor of 1.0078 when operated at the calibration voltage of + 400 V. In terms of beam quality correction factors for megavoltage beams, the in-house IC was found to exhibit characteristics similar to those of Scanditronix-Wellhofer IC 70 Farmer type IC. Conclusion: The constructed in-house Farmer-type IC was able to meet all the recommended characteristics for an IC, and therefore, the in-house IC is suitable for beam output calibration in external beam radiotherapy.en_US
dc.identifier.issn1898-0309
dc.identifier.otherdoi: 10.2478/pjmpe-2022-0022
dc.identifier.urihttp://ugspace.ug.edu.gh:8080/handle/123456789/38487
dc.language.isoenen_US
dc.publisherPolish Journal of Medical Physics and Engineeringen_US
dc.subjectin-house ionization chamberen_US
dc.subjectbeam output calibrationen_US
dc.subjectPerspexen_US
dc.subjectstem effecten_US
dc.subjectleakage currenten_US
dc.titleConstruction and pre-evaluation of an in-house cylindrical ionization chamber fabricated from locally available materialen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Construction-and-preevaluation-of-an-inhouse-cylindrical-ionization-chamber-fabricated-from-locally-available-materialsPolish-Journal-of-Medical-Physics-and-Engineering.pdf
Size:
834.27 KB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: