Evaluation and Validation of Computed Tomography Dose Accuracy (Ctdiw and Ctdivol)

dc.contributor.advisorKyere, A.K.
dc.contributor.advisorSosu, K.E.
dc.contributor.authorSintim, A.A.
dc.contributor.otherUniversity of Ghana, College of Basic and Applied Sciences, Department of Medical Physics
dc.date.accessioned2017-01-30T09:46:44Z
dc.date.accessioned2017-10-13T17:44:13Z
dc.date.available2017-01-30T09:46:44Z
dc.date.available2017-10-13T17:44:13Z
dc.date.issued2016-07
dc.descriptionThesis (MPhil) - University of Ghana, 2016
dc.description.abstractWeighted and average dose within a scan volume of a phantom have been evaluated and validated using two different devices and techniques. The Barracuda electrometer and Ion Chamber techniques were applied on a 16 slice Siemens CT scanner and the results compared to the console displayed CTDIw and CTDIvol valuesfor accuracy and compared to each other for validation purposes. With fixed exposure parameter of 130kVp and varying tube current-time products from 140mAs to 300mAs for the CT head phantom examination, there were varying deviations in both the CTDIw and CTDIvol from the two techniques. Tube currents of 140 mAs, 240 mAs and 300 mAs yielded 3.5%, 0.61% and -6.45% deviations when the respective CTDIvol values for both techniques were compared. There were mean CTDIvol of (42.3 + 8.6) mGy and (42.1 + 8.1) mGy for Barracuda and Ion Chamber techniques respectively with an average deviation of 1.4 mGy between them, when the tube currenttime products were varied from 140 – 300 mAs for the head phantom examination. Tube current-time products ranging from 80mAs – 220mAs were used for the CT body phantom examination and mean CTDIw measured were (16.6 + 6.7) mGy and (16.5 + 7.7) mGy for Barracuda and Ion Chamber techniques respectively with an average deviation of 1.0 mGy between them. The estimated head and body dose deviations were compared to international dose accuracy range. The results of the study showed that the deviations from the techniques were within a range of CTDIw and CTDIvol values which were favorably comparable to other similar retrospective research works, thus, the Ion Chamber technique can be used in place of the technique currently in use.en_US
dc.format.extentxv, 75p. ill.
dc.identifier.urihttp://197.255.68.203/handle/123456789/21452
dc.language.isoenen_US
dc.publisherUniversity of Ghanaen_US
dc.rights.holderUniversity of Ghana
dc.titleEvaluation and Validation of Computed Tomography Dose Accuracy (Ctdiw and Ctdivol)en_US
dc.typeThesisen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Evaluation and Validation of Computed Tomography Dose Accuracy (Ctdiw And Ctdivol) _ 2016.pdf
Size:
4.78 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 2 of 2
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:
Loading...
Thumbnail Image
Name:
license.txt
Size:
0 B
Format:
Item-specific license agreed upon to submission
Description: