Development and Verification of an Excel Program for Calculation of Monitor Units for Tangential Breast Irradiation with External Photon Beams

Abstract

The accuracy of MU calculation performed with Prowess Panther TPS (for Co-60) and Oncentra (for 6MV and 15MV x-rays) for tangential breast irradiation was evaluated with measurements made in an anthropomorphic phantom using calibrated Gafchromic EBT2 films. Excel programme which takes in to account external body surface irregularity of an intact breast or chest wall (hence absence of full scatter condition) using Clarkson’s sector summation technique was developed. A single surface contour of the patient obtained in a transverse plane containing the MU calculation point was required for effective implementation of the programme. The outputs of the Excel programme were validated with the respective outputs from the 3D treatment planning systems. The variations between the measured point doses and their calculated counterparts by the TPSs were within the range of -4.74% to 4.52% (mean of -1.33% and SD of 2.69) for the prowess panther TPS and -4.42% to 3.14% (mean of -1.47% and SD of -3.95) for the Oncentra TPS. The observed degree of deviation may be attributed to limitations of the dose calculation algorithm with in the TPSs, set up inaccuracies of the phantom during irradiation and inherent uncertainties associated with radiochromic film dosimetry. The percentage deviations between MUs calculated with the two TPSs and the Excel program were within the range of -3.45% and 3.82% (mean of 0.83% and SD of 2.25). The observed percentage deviations are within the 4% action level recommended by TG-114. This indicates that the Excel program can be confidently employed for calculation of MUs for 2D planned tangential breast irradiations or to independently verify MUs calculated with another calculation methods.

Description

Thesis (M Phil) University of Ghana - 2015
Thesis (MPhil) - University of Ghana,2015

Citation

Endorsement

Review

Supplemented By

Referenced By