Assessment of Mean Glandular Dose to Patients from Digital Mammography Systems

dc.contributor.advisorBoadu, M.
dc.contributor.advisorSchandorf, C.
dc.contributor.advisorSosu, E.
dc.contributor.authorPwamang, C.K.
dc.contributor.otherUniversity of Ghana, College of Basic and Applied Sciences, Department of Medical Physics
dc.date.accessioned2017-01-17T10:41:16Z
dc.date.accessioned2017-10-13T17:44:21Z
dc.date.available2017-01-17T10:41:16Z
dc.date.available2017-10-13T17:44:21Z
dc.date.issued2016-07
dc.descriptionThesis(MPHIL)-University of Ghana, 2016
dc.description.abstractMean glandular dose assessment of patients undergoing digital mammography examination has been done. A total of 297 patient data was used for the study. Basic Quality Control tests were done to ascertain the performance of the equipment used. The results of Quality Control tests indicated that the three Mammography units used for this study were functioning within the internationally acceptable performance criteria. Patients with a breast thickness of 30 mm within the two age groups of .40-49 yrs and 50-64 yrs received doses slightly higher than the acceptable dose levels. A patient in the category 40-49 yrs with breast thickness of 30 mm received 1.83 mGy as calculated Mean Glandular Dose, 2.10 mGy was the recorded dose and 1.58 mGy was recorded under the age group 50-64 yrs. These values have deviated by -22%, -40%, and -5.33% respectively from 1.5 mGy which is the recommended dose for a breast thickness of 30 mm. Also patients with breast thickness of 70 mm under the age group 40–49 yrs had a recorded dose of 6.58 mGy, which deviated by -1.21% from the recommended value of 6.5 mGy for that breast thickness. Aside these values, all the other values were within the recommended dose values. The percentage deviation between the recommended values and the calculated values was within ±25% which was a working limit that was set for this work. Doses delivered by the Full-field Digital mammography equipment were higher than doses delivered by the Computered Radiography equipment. The calculated Mean Glandular Doses for the three facilities were within recommended dose values.en_US
dc.format.extentXiii, 88p: ill
dc.identifier.urihttp://197.255.68.203/handle/123456789/21327
dc.language.isoenen_US
dc.publisherUniversity of Ghanaen_US
dc.rights.holderUniversity of Ghana
dc.subjectMean Glandular Doseen_US
dc.subjectDigital Mammography Systemsen_US
dc.subjectDigital Mammography Systemsen_US
dc.titleAssessment of Mean Glandular Dose to Patients from Digital Mammography Systemsen_US
dc.typeThesisen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Assessment of Mean Glandular Dose to Patients from Digital _July 2016.pdf
Size:
5.36 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 2 of 2
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:
Loading...
Thumbnail Image
Name:
license.txt
Size:
0 B
Format:
Item-specific license agreed upon to submission
Description: