Three-dimensional image quality test phantom for planar X-ray imaging

dc.contributor.authorNoonoo, J.B.
dc.contributor.authorSosu, E.
dc.contributor.authorHasford, F.
dc.date.accessioned2023-09-26T09:08:48Z
dc.date.available2023-09-26T09:08:48Z
dc.date.issued2023
dc.descriptionResearch Articleen_US
dc.description.abstractWe aimed to produce a simple, inexpensive 3D printed phantom as a prototype for image quality assessment of contrast, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR) and resolution in planar X-ray imaging systems. The test phantom was designed using SOLIDWORKS software, printed with a polylactic acid material and filled with paraffin wax. Circular aluminium sheets were used as inserts for contrast evaluation. A planar X-ray system was used for imaging and DICOM images were evaluated using ImageJ software. We evaluated spatial resolution, contrast, CNR and SNR. For resolution, full width at half maximum (FWHM) was measured on different grid sizes. For contrast, intensity of grey values and standard deviation were calculated on the different grid sizes. For CNR and SNR, difference in greyscale of investigated tissue and background per standard deviation of noise in the background was calculated. Resolution of the system was evaluated to be 1.57 and 1.80 lp/mm on grids A and B respectively. Contrast showed differential attenuation per variation in thickness. CNR increased from −13.7 for a thickness of 0.00 mm to 24.90 for a thickness of 28 mm. CNR did not change for a thickness greater than 16.0 mm. The SNR of the system fell in the acceptable range of ≥ 5. The results from the analyses performed indicate that the test phantom has great potential to be a good substitute for the commercially available phantoms on the market, especially for low-resource settings. Significance: This study highlights the emergence of 3D printing technology and its suitability within radiology and medical physics for the production of cost-effective phantoms which can serve as substitutes for commercial phantoms in low-resourced medical facilities in low- and middle-income countries.en_US
dc.identifier.otherdoi.org/10.17159/sajs.2023/14269
dc.identifier.urihttp://ugspace.ug.edu.gh:8080/handle/123456789/40114
dc.language.isoenen_US
dc.publisherVolume 119en_US
dc.subject3D printingen_US
dc.subjectphantomen_US
dc.subjectimage qualityen_US
dc.subjectcontrasten_US
dc.subjectresolutionen_US
dc.titleThree-dimensional image quality test phantom for planar X-ray imagingen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Threedimensional-image-quality-test-phantom-for-planar-Xray-imagingSouth-African-Journal-of-Science.pdf
Size:
1.05 MB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: