Patient skin dose measurement and risk of deterministic effect during fluoroscopy cardiac procedures

dc.contributor.authorInkoom, S.
dc.contributor.authorHasford, F.
dc.contributor.authoret al.
dc.contributor.authorNana, R.N.
dc.date.accessioned2023-12-20T11:28:14Z
dc.date.available2023-12-20T11:28:14Z
dc.date.issued2023
dc.descriptionResearch Articleen_US
dc.description.abstractThis study aimed at assessing patient’s peak skin doses (PSD) during fluoroscopy cardiac procedures and proposed a look-up table to enhance the patient’s dose management. Perspex phantom and thermoluminescent dosemeters (TLD) were irradiated for different dose levels with X-ray equipment (Philips Azurion 7). It was found that PSD measures were higher than the kerma at the interventional reference point [K (IRP)] reported with factors 1.55, 1.75 and 2.88 for anterior posterior (AP0o), left anterior oblique (LAO45o) and right anterior oblique (RAO45o), respectively. The equations describing the correlation between the PSD measured Kerma area product and cumulative air kerma were found with R-square values of 0.98 and 0.99, respectively. The statistical analysis shows a strong linear correlation between PSD and K (IRP) (P-value = 0.05). It was also found that 27% of the patients population considered in this work received a skin dose higher than the threshold of deterministic effect of 2 Gy and a look-up table with the equation of fitness was proposed to be implemented in the facility for K (IRP) higher than 500 mGy.en_US
dc.identifier.otherhttps://doi.org/10.1093/rpd/ncad206
dc.identifier.urihttp://ugspace.ug.edu.gh:8080/handle/123456789/41067
dc.language.isoenen_US
dc.publisherOxford University Pressen_US
dc.subjectpatient’s peak skin dosesen_US
dc.subjectmeasurementen_US
dc.subjectfluoroscopyen_US
dc.titlePatient skin dose measurement and risk of deterministic effect during fluoroscopy cardiac proceduresen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Patient skin dose measurement and risk of.pdf
Size:
542.48 KB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: