Evaluation of organ and effective doses using anthropomorphic phantom: A comparison between experimental measurement and a commercial dose calculator

dc.contributor.authorAkyea-Larbi, K.O.
dc.contributor.authorHasford, F.
dc.contributor.authorInkoom, S.
dc.contributor.authorTetteh, M.A.
dc.contributor.authorGyekye, P.K.
dc.date.accessioned2023-11-06T10:42:14Z
dc.date.available2023-11-06T10:42:14Z
dc.date.issued2023
dc.descriptionResearch Articleen_US
dc.description.abstractIntroduction: The aim of this study was to experimentally measure organ doses for computed tomog raphy (CT) procedures using thermoluminescence dosimeters (TLDs) on a RANDO anthropomorphic phantom and verify the measured doses using CT-Expo software. Methods: The phantom was irradiated using clinical CT scan protocols routinely used for specific pro cedures in the radiology department. Fifty TLD chips were used in this study. The scanning parameters (kVp, mA, s) used to scan the phantom were used as input parameters for CT-Expo dose estimations. Results: The TLD measured organ doses varied between 3.97 mGy for the esophagus and 56.22 mGy for the brain. High doses were recorded in the brain (37.80e56.22 mGy) and the eye lens (29.94 e36.16 mGy). Comparing the organ dose measurements between TLD and CT-Expo, the maximum organ dose difference was obtained for the eye lens. A comparison between the two methods for the other organs were all less than 32 %. The effective doses from the TLD measurements for the head, chest, and abdominopelvic CT examinations were 2.78, 6.67, and 17 mSv, respectively and CT-Expo were 2.20, 10.30, and 16.70 mSv, respectively. Conclusion: The experimental and computational results are comparable, and the reliability of the TLD measurements and CT-Expo dose calculator has been proven. Implications for study: A reason for the difference in dose measurements between the two methods has been attributed to the dissimilarity in the organ position in the Rando anthropomorphic phantom and the standard mathematical phantom used by CT-Expo. The experimental and computational results have been found to be comparable.en_US
dc.identifier.otherhttps://doi.org/10.1016/j.radi.2023.10.003
dc.identifier.urihttp://ugspace.ug.edu.gh:8080/handle/123456789/40685
dc.language.isoenen_US
dc.publisherRadiographyen_US
dc.subjectComputed tomographyen_US
dc.subjectOrgan doseen_US
dc.subjectEffective doseen_US
dc.subjectTLDen_US
dc.subjectAnthropomorphic phantomsen_US
dc.titleEvaluation of organ and effective doses using anthropomorphic phantom: A comparison between experimental measurement and a commercial dose calculatoren_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Evaluation of organ and effective doses using anthropomorphic.pdf
Size:
548.66 KB
Format:
Adobe Portable Document Format
Description:

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: