Development of Electron Density Phantom for Computed Tomography Image Quality Control Procedures
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
University of Ghana
Abstract
Computed tomography (CT) scanners are imaging technologies that generate high-resolution
cross-sectional pictures of anatomical structures with enhanced radiographic contrast. Ensuring
high-quality images is crucial for precise diagnostics in medical imaging settings of healthcare
institutions by utilising suitable phantoms for quality control protocols. The aim of the research is
to develop an electron density phantom for computed tomography image quality protocols. Melt
blending and geometric dilution procedures were employed to synthesise distinct tissue-mimicking
materials (TMMs) based on physical property (mass density), utilising weight compositions of
locally sourced test materials. The mass density of the TMMs was determined. The electron
density phantom was designed to be filled with water and accommodate TMM inserts and 3D
printed using polylactic acid filament. The phantoms with the inserted TMMs were scanned using
GE CT technology at 120 kVp and 200 mA, and the resultant CT images were analysed for image
quality evaluation with ImageJ software to identify human tissue based on literature studies. The
phantom was subsequently scanned using GE and Philips CT technology following medical
imaging protocols for the head and abdomen. It was found that the FTIR analysis of the TMMs
had bands that verified the presence of the test materials post-synthesis. The measured mean CT
numbers and standard noise of TMMs effectively mimicked several human tissues, including lung
of -547.39±14.61 HU, adipose of -73.28±9.60, kidney of 38.71±10.57 HU, muscle of 48.33±11.12
HU, liver of 67.36±9.94 HU, cortical of 1190.16±22.17 HU, and two trabecular bones of
202.05±14.01 HU and 446.75±18.60 HU in comparison to the HU range of human tissues from
existing literature works with noise of the TMMs below the 20 HU threshold. The linearity of the
CT numbers of TMMs, when scanned using the head and abdomen protocols on the two CT
machines, exhibited a strong positive correlation with R2 values of 0.99 and 0.93, respectively,and demonstrated high agreement of the CT numbers at p < 0.05. The results of this research
illustrate the precision of the developed electron density phantom, produced through 3D printing
with synthesised tissue-mimicking materials, which may be utilised for quality control procedures
in medical imaging settings.