Racial Disparities In Pulse Oximetry: Measurement Bias And Clinical Implications—A Systematic Review
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Frontiers in Anesthesiology
Abstract
This review synthesised existing evidence on racial disparities in pulse oximeter
accuracy, quantified the degree of bias, evaluated clinical effects, highlighted
implications for practice, and identified research priorities needed to achieve
equitable device performance. While prior systematic reviews have examined
pulse oximeter accuracy across varying skin pigmentation levels, most have
focused primarily on aggregate accuracy metrics without integrating
downstream clinical consequences or anaesthesiology-specific implications.
In contrast, this review synthesises paired SpO₂–SaO₂ comparisons stratified
by race or skin tone and explicitly links measurement bias to occult
hypoxaemia, treatment escalation thresholds, and perioperative decision-
making. By integrating device-level performance characteristics with clinical
risk
pathways, this review aims to provide actionable insights for
anaesthesiologists and critical care clinicians.
Description
Research Article
Citation
Ekor OE, Akiseku AK, Brocke SN-N and Ankamah S (2026) Racial disparities in pulse oximetry: measurement bias and clinical implications—a systematic review. Front. Anesthesiol. 5:1796674. doi: 10.3389/fanes.2026.1796674