Racial Disparities In Pulse Oximetry: Measurement Bias And Clinical Implications—A Systematic Review

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Frontiers in Anesthesiology

Abstract

This review synthesised existing evidence on racial disparities in pulse oximeter accuracy, quantified the degree of bias, evaluated clinical effects, highlighted implications for practice, and identified research priorities needed to achieve equitable device performance. While prior systematic reviews have examined pulse oximeter accuracy across varying skin pigmentation levels, most have focused primarily on aggregate accuracy metrics without integrating downstream clinical consequences or anaesthesiology-specific implications. In contrast, this review synthesises paired SpO₂–SaO₂ comparisons stratified by race or skin tone and explicitly links measurement bias to occult hypoxaemia, treatment escalation thresholds, and perioperative decision- making. By integrating device-level performance characteristics with clinical risk pathways, this review aims to provide actionable insights for anaesthesiologists and critical care clinicians.

Description

Research Article

Citation

Ekor OE, Akiseku AK, Brocke SN-N and Ankamah S (2026) Racial disparities in pulse oximetry: measurement bias and clinical implications—a systematic review. Front. Anesthesiol. 5:1796674. doi: 10.3389/fanes.2026.1796674

Endorsement

Review

Supplemented By

Referenced By